Аnalysis of the technical and dosimetric characteristics of the brachytherapy afterloaders. Ring applicator

https://doi.org/10.35596/1729-7648-2021-19-5-13-20

Abstract

A study of the technical and dosimetry characteristics of brachytherapy afterloaders and applicators was carried out. Ring applicator has been taken as an example, the correctness of positioning of a radiation source (RS) inside the applicators was tested as part of the comissioning and quality control procedure of the brachytherapy applicators. The magnitudes of inconsistencies in the position of RS were established when planning and implementing treatment plans for radiation therapy. The identification of the values of the discrepancy was carried out using the obtained X-ray images of the applicator at the time of the implementation of the irradiation plans. The treatment plan was a sequential positioning of RS in the body of the applicator in each active position with a minimum step from the tip to the vaginal part of the applicator. The X-ray image was obtained by locating the source sequentially at each active position of the applicator. When carrying out dosimetric planning, 3 methods of applicator reconstruction were used. The analysis revealed that the applicator reconstruction method affects the magnitude of the discrepancy in determining the position of the source in the lumen of the applicator ring. Using the methods of statistical analysis, the mean, median, maximum and minimum values of the detected deviations were calculated. The results are presented in the form of tables and graphs for all investigated stop positions of IRS. Based on the results of the study, we consider it expedient to carry out quality control procedures when putting applicators into clinical operation. Based on the results obtained, we consider it acceptable to conduct a quality control procedure for the positioning accuracy of radiation sources in the applicators at least 1 time per month. Taking into account the results of the study when carrying out dosimetric planning will improve the quality of the irradiation sessions using the brachytherapy method, thereby improving the quality of oncological care for the population.

About the Authors

D. I. Kazlouski
International Sakharov Environmental Institute of Belorussian State University
Belarus

Kazlouski Dzianis I., Postgraduate student

220070, Minsk, Dolgobrodskaya str., 23/1



Y. V. Titovich
ADANI
Belarus

Titovich Yahor V., PhD, Medical Physicist

Minsk



Y. I. Kazlouskaya
N.N. Alexandrov National Cancer Centre
Belarus

Kazlouskaya Yuliya I., Medical Physicist

Minsk



References

1. Fonseca G., R van den Bosch M., Voncken R., Podesta M. A novel system for commissioning brachytherapy applicators: Example of a ring applicator. Physics in Medicine and Biology; 2017.

2. McMahon R., Zhuang T., Beverly A Steffey, Song H. Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets. Journal of applied clinical medical physics. 2011;12(4).

3. Hoshyar N., Cohen G., Voros L., Zaider M. Commissioning of Tandem and Split Ring Applicator. Brachytherapy. 2019; May 01;15(1):S153.

4. Tanderup K., Hellebustb T.P., Langc S., Granfeldtd J., Pötterc R., Lindegaarda J.C., Kirisitsc C. Consequences of Random and Systematic Reconstruction Uncertainties in 3D Image Based Brachytherapy in Cervical Cancer. Radiotherapy and oncology: journal of the ESTRO. 2008;Nov.:56-63.

5. Awunor O., Dixon B., Walker C. Direct reconstruction and associated uncertainties of (192)Ir source dwell positions in ring applicators using gafchromic film in the treatment planning of HDR brachytherapy cervix patients. Phys. Med. Biol. 2013;58:3207-3225.

6. Tam C., Wronski M., Au P., Ravi Ananth A. Medical Physics, Sunnybrook Odette. EPID-based quality assurance technique for HDR ring and tandem source dwell positions with the Nucletron flexitron afterloader. Brachytherapy. 2014;13:S15eS126.

7. Stern R., Liu T. Dwell position inaccuracy in the Varian GammaMed HDR ring applicator. Journal of Applied Clinical Medical Physics. 2010;11(4):3158.

8. Gerald J. Kutcher, TG Chair, Lawrence Coia, Michael Gillin, William F. Hanson, Steven Leibel, Robert J. Morton, Jatinder R. Palta, James A. Purdy, Lawrence E. Reinstein, Goran K. Svensson, Mona Weller, Linda Wingfield. Comprehensive QA for radiation oncology: Report of AAPM radiation therapy committee task group 40. Med Phys. 1994;21:581-618.

9. Awunor O., Berger D., Kirisits C. A multicenter study to quantify systematic variations and associated uncertainties in source positioning with commonly used HDR afterloaders and ring applicators for the treatment of cervical carcinomas. Medical physics 2015;42(8):4472-4483.


Review

For citations:


Kazlouski D.I., Titovich Y.V., Kazlouskaya Y.I. Аnalysis of the technical and dosimetric characteristics of the brachytherapy afterloaders. Ring applicator. Doklady BGUIR. 2021;19(5):13-20. (In Russ.) https://doi.org/10.35596/1729-7648-2021-19-5-13-20

Views: 4101


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1729-7648 (Print)
ISSN 2708-0382 (Online)